Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Angus Electric POWER MASTER IIIB: (1 yr) CAL VER /5520 DATE: 26-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 9 NUMBER OF LINES: 111 CONFIGURATION: Fluke 5520A STANDARD: Fluke 5500A/COIL ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 STD Fluke 5500A/COIL 1.002 ASK- P F V 1.003 ASK+ K X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Turn UUT power switch to ON. 1.005 DISP 1.005 DISP Warm-up time: NONE 1.005 DISP Temperature: 23 +/-3 degC 1.005 DISP Humidity: 55 +/-10% 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 HEAD {AC VOLTAGE} 1.008 DISP Connect the 5520A and the UUT as follows: 1.008 DISP [32] 5520A NORMAL HI to UUT Red Voltage Input 1.008 DISP [32] 5520A NORMAL LO to UUT White Voltage Input 1.009 DISP Set Function to Volts 1.009 DISP Set Element to RD 1.009 DISP Set Monitor to Center 1.009 DISP Set Mode to Run 1.010 5520 60.0V 1% 0.1U 60H SI 2W 2.001 5520 600V 1% 1U 60H SI 2W 3.001 HEAD {AC CURRENT} 3.002 DISP Disconnect the voltage leads from the UUT. 3.002 DISP 3.003 DISP Connect the STANDARD CURRENT TRANSDUCER(clamp) that is 3.003 DISP supplied with the UUT, to the POWER MASTER IIIB. 3.004 DISP Connect the 5500A/COIL to the 5520A 20A and AUX LO 3.004 DISP terminals. 3.004 DISP 3.004 DISP Clamp the CURRENT TRANSDUCER around the coil. 3.005 DISP Set Function to Amps 3.006 5520 500A 1% 1U 60H SI HC T5 4.001 DISP Connect a 5500A/COIL to the 5520A AUX HI and LO 4.001 DISP terminals. 4.001 DISP 4.001 DISP Clamp the CURRENT TRANSDUCER around the coil. 4.002 5520 50.0A 1% 0.1U 60H SI T5 5.001 HEAD {POWER} 5.002 DISP Leave the CURRENT TRANSDUCER connected. 5.002 DISP 5.002 DISP Re-connect the voltage leads: 5.002 DISP [32] 5520A NORMAL HI to UUT Red Voltage Input 5.002 DISP [32] 5520A NORMAL LO to UUT White Voltage Input 5.003 DISP Set Function to kWatts 5.004 M5520 100V SI 5.005 5520 50A 60H SI S T5 5.006 ACC 5.00kW 0.0466U 5.007 MEMI Enter the UUT Power reading kilowatts. 5.008 MEME 5.009 MEMC 5.00kW 0.1U 6.001 DISP Set Function to PF%/AVG 6.002 M5520 100V SI 6.003 5520 5kW 60H SI S T5 6.004 ACC 100.0dpf 0.00% 6.005 MEMI Enter UUT reading in DPF. 6.006 MEME 6.007 MEMC dpf 0.6% 7.001 M5520 * 7.002 DISP Disconnect the 5500A/COIL from the 5520A AUX terminals. 7.003 HEAD {PHASE ANGLE} 7.004 DISP Connect the 5520A to the UUT as follows: 7.004 DISP [32] NORMAL HI to UUT Red Voltage Input 7.004 DISP [32] NORMAL HI to UUT Blue Voltage Input 7.004 DISP [32] NORMAL LO to UUT White Voltage Input 7.005 DISP Set Function to Volts 7.005 DISP Set ELEMENT to T/angle 7.006 5520 60V 60H SI S 2W 7.007 MEMI Enter UUT reading in degrees. 7.008 MATH MEM1 = MEM - 180 7.009 JMPT 8.002 7.010 ACC 0deg 0.15U 7.011 MEME 7.012 MEMC deg 3U 8.001 JMP 9.001 8.002 ACC 360deg 0.15U 8.003 MEME 8.004 MEMC deg -3U 9.001 END